Silver/Sulfide Half-Cell Ion Selective Electrode (ISE) - HI4015
- Silver detection from 0.11 to 107,900 mg/L Ag
- Sulfide detection from 0.003 to 32,100 mg/L S2-
- For use with HI5315 reference half-cell
Silver (Ag )
For the HI4015 to measure silver accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the halide ionic strength adjuster (ISA) solution. The ISA ensures that the activity coefficient is constant so the concentration of free silver ions can be measured. Conversion Table for AgMultiply | |
moles/L (M) to ppm (mg/L) |
1.0787 x 105 |
ppm (mg/L) to M (moles/L) |
9.27 x 10-6 |
Sulfide (S2-)
For accurate sulfide measurement, it is important that sulfur is released from hydrogen and that sulfide oxidation is prevented with the sulfide antioxidant buffer solution (SAOB). The SAOB ensures that the solution pH is above pH 12, reducing the effect of interfering ions and releasing sulfur from hydrogen ions. SAOB also acts to stabilize the activity coefficient so the concentration of free sulfide ions can be measured. Conversion Table for S2-Multiply | |
moles/L (M) to ppm (mg/L) |
3.207 x 104 |
ppm (mg/L) to M (moles/L) |
3.12 x 10-5 |
Features at-a-glance
Fixed Sensing Modules
- Solid-state half-cell ISEs are electrodes that detect free ions in a solution. The highly insoluble, compressed sensing pellet is attached to the HI4015 silver/sulfide half-cell by a fixed sensing module for ease of use.
Durable Epoxy Body
- The probe body of the silver/sulfide half-cell ISE is composed of durable epoxy resin.
BNC Connection
- The HI4015 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.
Theory of Operation
A solid-state sensor electrode develops a voltage due to the ion exchange that occurs between the solution and the inorganic membrane. An equilibrium occurs due to the highly limited solubility of the membrane material in the sample. When the ionic strength of the solution is fixed by the addition of ISA, the voltage is proportional to the concentration of free ions in solution organic ion exchange resin. When the ionic strength of the solution is fixed by the addition of ISA, the voltage is proportional to the concentration of free ions in solution.Specification
SKU | HI4015 |
Type | solid state; half-cell |
Measurement Range | 1.0M to 1•10-6M 107900 to 0.11ppm (Ag ) 1.0M to 1•10-7M 32100 to 0.003 ppm (S2-) |
Optimum pH Range | 2 to 8 (Ag ) 12 to 14 (S2-) |
Temperature Range | 0 to 80°C |
Diameter | 12 mm |
Length | 120 mm |
Body Material | epoxy |
Cable | 1 m coaxial |
Common Applications | used as an indicator for titrations using silver nitrate; for the determination of sulfide ions in waters, paper liquors, natural waters and soils. |
Connection | BNC |