Portable Photometer for Bromine, Free and Total Chlorine, Cyanuric Acid, Iron LR, Iodine and pH - HI97101
• CAL Check™ · Validate instrument performance at any time using CAL Check cuvettes made with NIST traceable standards. The CAL Check screen guides the user step-by-step through the validation process and user calibration. • On-screen tutorial mode with animations · Guides users step-by-step through the measurement process • Waterproof and floating IP67 case • Unit of measure is displayed along with reading • Built-in timer · Built-in reaction timer that ensures consistency between tests. • Error messages on display · Alerts to problems including no cap, high zero, and standard too low • GLP data · Displays the last calibration date. • Auto logging • Battery status indicator • Auto-shut off
Specifications
pH | Range | 6.5 to 8.5 pH | |
Resolution | 0.1 pH | ||
Accuracy @25°C (77°F) | ±0.1 pH | ||
Method | Phenol Red method | ||
Chlorine ,Free and Total | Range | 0.00 to 5.00 mg/L (ppm) | |
Resolution | 0.01 mg/L | ||
Accuracy @25°C (77°F) | ±0.03 mg/L ±3% of reading | ||
Method | adaptation of the USEPA method and Standard Method 4500-Cl G | ||
Cyanuric Acid | Range (all methods) | 0 to 80 mg/L (ppm) | |
Resolution (all methods) | 1 mg/L | ||
Accuracy @25°C (77°F) | ±1 mg/L ±15% of reading | ||
Method | adaptation of the turbidimetric method | ||
Iodine | Range | 0.0 to 12.5 mg/L (ppm) | |
Resolution | 0.1 mg/L | ||
Accuracy @25°C (77°F) | ±0.1 mg/L ±5% of reading | ||
Method | adaptation of the EPA, DPD method | ||
Bromine | Range | 0.00 to 10.00 mg/L (ppm) | |
Resolution | 0.01 mg/L | ||
Accuracy @25°C (77°F) | ±0.08 mg/L ±3% of reading | ||
Method | adaptation of the EPA, DPD method | ||
Iron LR | Range | 0.00 to 1.60 mg/L (ppm) | |
Resolution | 0.01 mg/L | ||
Accuracy @25°C (77°F) | ±0.01 mg/L ±8% or reading | ||
Method | adaptation of the TPTZ method | ||
Measurement System | Light Source | light emitting diode | |
Bandpass filter | 525 nm | ||
Bandpass filter bandwidth | 8 nm | ||
Bandpass filter wavelength accuracy | ±1.0 nm | ||
Light Detector | silicon photocell | ||
Cuvette type | round 24.6 mm diameter (22 mm inside) | ||
Additional Specifications | Auto logging | 50 readings | |
Display | 128 x 64 pixel B/W LCD with backlight | ||
Auto-off | after 15 minutes of inactivity (30 minutes before a READ measurement) | ||
Battery type / Life | alkaline 1.5 V AA (3) / > 800 measurements (without backlight) | ||
Environment | 0 to 50°C (32 to 122°F); 0 to 100% RH, non-serviceable | ||
Dimensions | 142.5 x 102.5 x 50.5 mm (5.6 x 4.0 x 2.0") | ||
Weight | 380 g (13.4 oz.) | ||
Ordering Information | HI97101 is supplied with sample cuvettes (2), sample caps (2), plastic stoppers (2), 1.5V AA batteries (3), instrument quality certificate, and instruction manual.CAL Check standards and testing reagents sold separately HI97101C includes photometer, CAL Check standards, sample cuvettes (2), sample caps (2), plastic stoppers (2), 1.5V AA batteries (3), cuvette wiping cloth, CAL Check standard certificates, instrument quality certificate, instruction manual, and rigid carrying case.Reagents sold separately |