Check these for elements for proper pH measurements: meter, electrode, buffer, and sample.

pH Meter

  • Confirm the meter settings are correct.
  • Confirm that your temperature sensor is connected (if required).
  • Check that the batteries & power cord are in good working order.
  • Check that the electrode connection is secure.
  • Calibrate frequently to at least 2 points with fresh buffers.

Replace if the meter does not turn on or shows signs of severe wear.

Always remember to reference your manual

Premium Content

Login to your account or create one right now to access to our exclusive content.

Electrode

  • Inspect your electrode cord and connection for damage.
  • Check that the meter connection is secure.
  • Inspect the pH bulb for scratches.
  • Confirm the pH bulb isn’t broken.
  • Use cleaning solution to ensure electrode bulb & junction are clean.
  • Use storage solution to ensure the electrode is hydrated.
  • Use the correct electrode type for sample being tested.

Replace if electrode won’t stabilise, calibrate or is broken or scratched.

Place in storage solution at least 1 hour or overnight to fully hydrate

pH Buffers

  • Stir buffer with magnetic stirrer or probe.
  • Confirm you are using the correct pH calibration buffer.
  • Inspect your buffers for visible contamination.
  • Use fresh buffer for calibration.
  • Confirm buffer is not expired or bottle has not been opened for more than 4-8 weeks; or 1-2 weeks for buffer > pH 7.

Replace buffers if contamination is present, they are expired or have been opened for too long.

You should “bracket” your expected result e.g.: if the expected result = pH 5, use buffers pH 4.01 and pH 7.01.

Sample

  • Stir sample with magnetic stirrer or probe.
  • Confirm you have enough sample to cover the electrode bulb and junction.
  • Confirm the temperature of your sample will not damage electrode.
  • Sample is appropriate for probe type
  • Sample is fresh and representative of the process/product.

Clean Regularly | Calibrate Often | Condition Always

 Download your free copy of the “pH Measurements Checklist” here.