SKU
HI4013
Nitrate Half-Cell Ion Selective Electrode (ISE) - HI4013
€937.00
In stock
The HI4013 is a liquid membrane, half-cell ion selective electrode (ISE) for the determination of nitrate (NO3-) in solution. The PVC membrane comprised with organic ion exchanger produces a potential change due to the nitrate ion exchange between the organic polymer membrane and the sample. The internal sensing elements are housed within a durable epoxy body. The HI4013 is ideal for a variety of applications in the food industry, agriculture, and water analysis.
- Liquid membrane sensor
- Detection from 0.62 to 6,200 mg/L NO3-
- For use with HI5315 reference half-cell
Multiply | |
moles/L (M) to ppm (mg/L) | 62000 |
ppm (mg/L) to M (moles/L) | 1.61 x 10-5 |
Features at-a-glance
Replaceable Sensing Modules
- Half-cell liquid membrane ISEs are electrodes that detect free ions in a solution. This type of ISE is designed to incorporate easy replacement of the sensing module.
Durable Epoxy Body
- The probe body of the nitrate half-cell ISE is composed of durable epoxy resin.
BNC Connection
- The HI4013 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.
Theory of Operation
A liquid membrane ion selective electrode develops a voltage due to the ion exchange that occurs between the solution and the polymer membrane that contains organic ion exchange resin. When the ionic strength of the solution is fixed by the addition of ISA, the voltage is proportional to the concentration of free ions in solution.Specifications
SKU | HI4013 |
Type | polymer membrane; half cell |
Measurement Range | 1.0M to 1•10-5M 6200 to 0.62 mg/L (ppm) 1400 to 0.4 mg/L (ppm) as N |
Optimum pH Range | 3.0 to 8 |
Temperature Range | 0 to 40°C |
Diameter | 12 mm |
Length | 120 mm |
Body Material | epoxy/PVC |
Cable | 1 m coaxial |
Common Applications | determination of free nitrate in natural waters (fresh and sea), and in emulsified food and plant samples |
Connection | BNC |