Mini Titrator for Measuring Titratable Acidity in Fruit Juice - HI84532
- Complete kit for titratable acidity in fruit juice.
- Dynamic dosing makes testing both faster and more accurate.
- Great for small & medium wineries and advanced hobbyists.
Fruits, Juices | Titratable acidity (g/100 mL) | Predominant acid |
Apple, pear | 0.36-0.80 | Malic Acid |
Cranberry | 1.6-3.6 | Citric Acid |
Grapefruit | 1.2-2.0 | Citric Acid |
Lemon | 4-6.2 | Citric Acid |
Mango | 0.34-0.84 | Citric Acid |
Orange | 0.8-1.4 | Citric Acid |
Peach, nectarine, sweet cherry | 0.24-0.94 | Citric Acid |
Pineapple | 0.7-1.6 | Citric Acid |
Plum, sour cherry | 0.94-1.64 | Malic Acid |
Strawberry | 0.6-1.1 | Citric Acid |
Table Grape | 0.4-0.9 | Tartaric Acid |
Tomato | 0.34-1.00 | Citric Acid |
Double Junction pH Electrode
The HI84532U-01 is supplied with the HI1131B refillable, double junction, combination pH electrode. By design, the HI1131B has a spherical tip for use in aqueous or liquid solutions. This versatile electrode provides a wide surface of contact with a sample and is ideal for direct measurements or titrations in the beverage industry.Custom pH Calibration
According to AOAC International, the titration endpoint for fruit products is pH 8.1. The HI84532U-01 uses a custom calibration point of pH 8.20. The preprogrammed pH 4.01 and pH 8.20 calibration points allow for calibration of the meter that brackets the pH readings.Good Laboratory Practice
The HI84532U-01 offers comprehensive GLP information including date and time of the last pH electrode and dosing pump calibrations. Tracking calibrations are critical to have confidence in the results obtained from titrations. The GLP data is stored along with any logged readings.Piston Driven Dosing Pump
The heart of the HI84532U-01 is the piston-driven burette pump. This type of dosing system uses a motor in which each dose is very precisely controlled and the volume dispensed is accurately determined. This titrant is dynamically dosed, where titrant is delivered in larger doses at the start of the titration and smaller doses near the endpoint. With larger doses at the beginning of the titration, the speed of the titration is increased, where smaller doses near the endpoint allow for more time for the titrant and analyte to react. Smaller doses also prevent the over titration of a sample and a more accurate determination of titrant volume used.Automatic Stirrer
The built-in stirrer is automatically maintained at a speed of 600 rpm, regardless of the viscosity of the solution being titrated.On-Screen Features
Easy-to-Use Interface
The HI84532U-01 has an intuitive user design with clearly defined keys and a large screen that is easy to navigate. The meter has a built-in tutorial mode that, when enabled, will walk the user step by step through the titration process. A dedicated HELP key is always available to allow access to content-specific information during both calibration and titration.
Procedure Warnings
Users are warned if there is an error in procedures such as the titration exceeded the maximum volume of titrant. Graphic Mode This mini titrator displays in-depth data during titration, including a real-time graph of the titration curve.
This mini titrator displays in-depth data during titration, including a real-time graph of the titration curve.
CAL Check™
Accuracy is always ensured with Hanna’s unique CAL CheckTM feature, which analyzes the response of the electrode during the calibration process. Based on the electrode response in the buffer, indicators are displayed on the screen to alert the user of potential problems during calibration. These indicators include Buffer Contaminated, Electrode Dirty/Broken, and overall probe condition. The CAL CheckTM function not only ensures an accurate pH reading when the HI84532U-01 is used as a pH meter but also an accurate titration since the endpoint of a titratable acidity titration is determined by a set pH value.
Log-on-Demand
The HI84532U-01 allows for data logging of up to 400 samples: 200 titration results and 200 pH/mV readings. Data can be stored and exported to a USB drive or a PC using the USB connection.
pH/mV Meter
In addition to being an automatic titrator, the HI84532U-01 also can be used as a pH/mV meter. As a pH meter, it has many features of a professional-grade benchtop including automatic calibration up to three points with four available buffers, a 0.01 pH resolution, an accuracy of ±0.01 pH, automatic temperature compensation and comprehensive GLP data.
Specifications:
Specification Name | Detail |
---|---|
sku | HI84532U-01 |
Range | Low Range: g/100 mL as citric acid: 0.10 to 2.00% CA; g/100 mL as tartaric acid: 0.11 to 2.35% TA; g/100 mL as malic acid: 0.10 to 2.09% MA High Range: g/100 mL as citric acid: 1.00 to 10.00% CA; g/100 mL as tartaric acid: 1.17 to 11.72% TA; g/100 mL as malic acid: 1.05 to 10.47% MA |
Resolution | 0.01% |
Accuracy (@25ºC/77ºF) | ±0.02% or 3% of reading CA whichever is greater |
Sample Volume | 5 mL |
Methods | acid base titration |
Principle | endpoint titration: 8.1 pH |
Pump Speed | 10 mL/min |
Stirring Speed | 600 rpm |
pH Range | -2.0 to 16.0 pH; -2.00 to 16.00 pH |
pH Resolution | 0.1 pH / 0.01 pH |
pH Accuracy (@25ºC/77ºF) | ±0.01 pH |
pH Calibration | one, two, or three-point calibration, four available buffers (4.01, 7.01, 8.20, 10.01) |
Temperature Compensation | manual or automatic |
mV Range | -2000.0 to 2000.0 mV |
mV Resolution | 0.1 mV |
mV Accuracy (@25ºC/77ºF) | ±1.0 mV |
Temperature Range | -20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K |
Temperature Resolution | 0.1°C; 0.1°F; 0.1K |
Temperature Accuracy (@25ºC/77ºF) | ±0.4°C; ±0.8°F; ±0.4 K |
Logging Data | up to 400 ( 200 titrations, 200 pH/mV readings) |
Electrode | HI1131B glass body, refillable, with BNC connector and 1 m (3.3’) cable (included) |
Temperature Probe | HI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
Power Supply | 12 VDC adapter (included) |
Dimensions | 235 x 200 x 150 mm (9.2 x 7.9 x 5.9”) |
Weight | 1.9 kg (67.0 oz.) |
Ordering Information | HI84532 is supplied with HI1131B pH electrode, HI84532-50 low range titrant (100 mL), HI84532-51 high range titrant (100 mL), HI84532-55 calibration standard (100 mL), HI7662-T temperature probe, HI7082 electrode fill solution (30 mL), 100 mL beakers (2), 20 mL beaker, tube set (aspiration tube with titrant bottle cap and dispensing tube with tip), dosing pump valve, 5 mL syringe, 1 mL plastic pipette, stir bar, power adapter, instruction manual and quality certificate. |
Warranty | 2 years |