SKU
HI96803

Digital Refractometer for % Glucose by Weight Analysis - HI96803

€253.00
In stock
The HI96803 is a rugged, portable digital refractometer for food products designed to report the content of aqueous solutions containing sugar as % glucose by weight (% w/w). The HI96803 report the results with an accuracy of ±0.2%. The operation of the meter is simplified with only two buttons: one button is to calibrate with distilled or deionized water and the other to take a measurement. All readings are automatically compensated for temperature variations according to the ICUMSA Methods Book standard and displayed within a 1.5 second response time. The sealed flint glass prism and stainless steel well are easy to clean. Just wipe with a soft cloth in preparation for the next sample.

The HI96803 portable refractometer converts the refractive index of a food sample to % glucose by weight. This conversion is based on the tables found in the ICUMSA Methods Book (International Commission for Uniform Methods of Sugar Analysis) that documents the changes in refractive index with temperature for a percent by weight glucose solution. The HI96803 is ideal for the analysis of fruits, energy drinks, puddings, soy milk, juices, jam, marmalade, honey, soups, jelly, tofu and condiments. The HI96803 is an easy to use tool for measuring sugar in food samples in the field or in the lab.

  • Designed for food sugar analysis
  • Temperature Compensation algorithms based on glucose solution
  • 0 to 85% glucose by weight range with an accuracy of ± 0.2%

Making a % Standard Glucose Solution
To make a glucose Solution, follow the procedure below:

  • Place container (such as a glass vial or dropper bottle that has a cover) on an analytical balance.
  • Tare the balance.
  • Weigh out X grams of high purity glucose (CAS #: 50-99-7) directly into a beaker
  • Add distilled or deionized water to the beaker so the total weight of the solution is 100g.
  • Note: Solutions above 60% may need to be vigorously stirred or shaken and heated in a water bath to roughly 40°C (104°F). Remove solution when fructose has dissolved and allow to cool before use.

 

Refractive Index

The HI96803 takes measurements based on a sample’s refractive index. Refractive index is a measurement of how light behaves as it passes through the sample. Depending on the sample’s composition, light will refract and reflect differently. By measuring this activity with a linear image sensor, the sample’s refractive index can be assessed and used to determine its physical properties such as concentration and density. In addition to the linear image sensor, the HI96803 uses an LED light, prism and lens to make measurement possible.

Variations in temperature will affect the density of a solution based on the compound that is present. In digital refractometry the use of temperature compensation is necessary for accurate results. The HI96803 contains a built-in temperature sensor and is programmed with temperature compensation algorithms in accordance with the ICUMSA Methods Book for a percent by weight sucrose solution.

Features at-a-glance:


One-point calibration

  • Calibrate with distilled or deionized water


Small sample size

  • Sample size can be as small as 2 metric drops (100 μl)


Automatic Temperature Compensation (ATC)

  • Samples automatically compensated for temperature variations


Quick, accurate results

  • Readings are displayed in approximately 1.5 seconds


Dual-level LCD

  • The dual-level LCD displays measurement and temperature readings simultaneously
  • Stainless steel sample well
    Easy to clean and corrosion-resistant

  • Battery indicator
    Battery percent level remaining at startup and low battery indicator

  • Automatic shut-off
    To conserve battery life the meter shuts off automatically after three minutes of non-use

  • Stainless steel sample well
    Easy to clean and corrosion-resistant

  • IP65 water resistant protection
    Water resistant ABS plastic casing designed to perform under laboratory and field conditions.

 

Specifications

Sugar Content Range

0 to 85% Brix (% w/w glucose)

Resolution  

0.1% Brix mass

Accuracy 

±0.2% Brix mass

Temperature Range

0 to 80°C (32 to 176°F)

Resolution 

0.1°C (0.1°F)

Accuracy 

±0.3 °C (±0.5 °F)

Temperature Compensation

automatic between 10 and 40°C (50 to 104°F)

Minimum Sample Volume

100 μL (to cover prism totally)

Sample Cell

stainless steel ring and flint glass prism

Measurement Time

 approximately 1.5 seconds

Battery Type

9V / approximately 5000 readings

Auto-off

after three minutes of inactivity

Refractometer Light Source

yellow LED

Dimensions 

192 x 102 x 67 mm (7.6 x 4.01 x 2.6”)

Weight 

420 g (14.8 oz.)

Ordering information