Ammonia Standards (Low Range) CAL Check™ HI97700-11
Details
HI97700-11 is a high quality set of calibration standards that allows users to validate and calibrate the HI96700 Ammonia Low Range photometer. The HI97700-11 includes one sealed “A” cuvette for zeroing the meter and one “B” cuvette that is used to validate or calibrate. The standard value of the “B” cuvette is 1.50 ±0.01 mg/L (ppm) ammonia nitrogen (NH3-N). An acceptable reading when validating the HI96700 with the standards is 1.50 ±0.05 mg/L @ 25°C. The calibration set is produced in our state-of-the-art manufacturing facility using reagent grade chemicals in a temperature controlled environment.
CAL Check™ standards with certificate
The HI97700-11 CAL Check™standards are used for calibration and performance verification of HI977 photometers with the CAL Check function.
Supplied with Certificate of Analysis
- Lot number
- Expiration date
- Standard value @ 25 °C
- Reference meter NIST traceable
Provided storage containers
- Light tight
- Protects from accidental breakage
Specifications
Package | Sealed standards: 0.00 and 1.50 mg/L |
---|---|
Quantity | 1 set |